A:148 AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
2) Select the Tools tab.
3) Press Auto schedule.
4) Press Rinse/Disinfection process.
5) Select Heat.
6) Press OK.
7) Select day.
8) Set the time when the program should
finish.
9) Press OK.
10) To schedule more days in the week: repeat
steps 7-9.
11) To perform a heat disinfection with
CleanCart cartridge: Place a CleanCart
cartridge into the cartridge holder before
the disinfection starts.
10.3.7.3 To turn off a scheduled program
Procedure
1) Press the Disinfection button.
2) Select the Tools tab.