AK 98 Dialysis Machine - Attentions B:113
539 Fluid tube sensor test
539 Fluid tube sensor testī¼
Attach fluid tubes to safety couplings. To
start disinfection press Confirm.ī¼
Appears:
When the dialysis machine cannot detect that the dialysis
fluid tubes are attached to the safety couplings. The
corresponding machine test cannot be made.
Machine actions:
The ongoing disinfection or rinse program is paused.
Operator action:
To continue the disinfection or rinse program:
1. Check that the dialysis fluid tubes are properly attached
to the safety couplings.
2. Press Confirm to continue the disinfection or rinse
program.
As an alternative, consult an authorised technician.
Mute time:
2 minutes.