AK 98 Dialysis Machine - Attentions B:171
602 UF volume is achieved too early
602 UF volume is achieved too early
Check and adjust min UF rate set value. To
continue press Confirm.
Appears:
When the set UF volume has been achieved and the
remaining treatment time is more than 20 minutes.
Machine actions:
Alarm sound is delayed for 2 minutes. The machine
automatically sets the actual UF rate to minimum UF rate.
Operator action:
1. Press the Fluid button.
2. Select the UF rate tab.
3. Check that the minimum UF rate is suitable the patient.
If not, adjust the value. If necessary, consider adjusting
treatment time or UF volume within the limits for the
patient“s needs and tolerance.
Mute time:
2 minutes.