AK 98 Dialysis Machine - Attentions B:155
584 Priming volume limit is achieved
584 Priming volume limit is achieved
To continue press Confirm.
Appears:
When the extra priming volume limit has been reached.
The extra priming volume limit can be preset.
Machine actions:
None.
Operator action:
If this attention appears during initiation of the treatment,
check that the venous blood line is properly inserted into
the priming detector.
Press Confirm.
Mute time:
2 minutes.