A:110 AK 98 Dialysis Machine - Haemodialysis - Single needle treatment
HCEN12745 Revision 10.2016
Program version 2.xx
NOTE!
The blood pump is preset to
stop when the priming
detector detects blood in the
venous blood line and the
blood path in the flow
diagram lights up.
10) Press the flashing Blood pump button to
restart the blood pump.
11) Adjust the blood flow rate using the Blood
pump up and Blood pump down buttons.
5.4 Start the treatment
WARNING!
The operator must take proper precaution in order to prevent coagulation in the
extracorporeal circuit.
Coagulation may lead to:
ā inadequate delivery of dialysis.
ā risks associated with movement of blood clots to the patient.
ā disabling of the air detector function if blood clots aggregate in the drip chamber.
WARNING!
To avoid patient blood loss and damage to patient blood cells, always follow the
manufacturer's instructions for use regarding the dialyzer and blood lines. When
attaching the blood lines, ensure that all connections are properly secured and that
the blood lines are not kinked. If the heparin pump is not used, make sure that the
cap at the end of the thin line used for the heparin solution is closed properly.
CAUTION!
Check that the blood path of the flow diagram lights up. If it does not light up, check
that the venous blood line is correctly placed in the priming detector. If the blood
path does not light up, the machine considers the patient not connected and as a
consequence the UF system is not controlling the patient ultrafiltration. If the air
detector has not been activated during priming it will be automatically activated
when the priming volume is achieved.