A:168 AK 98 Dialysis Machine - IT Connectivity
HCEN12745 Revision 10.2016
Program version 2.xx
12.3 Cancel prescription retrieval
If Cancel is pressed the identified patient and the patient prescription will be discarded
and all set parameters will revert to default.
To retrieve the prescription again repeat the steps in Section 12.2 ā€œConfirmed Patient
ID and retrieval of patient prescription when the Patient ID is confirmedā€ on
pageī™›A:166.
A prescription can be retrieved until treatment is started or until the prescription is
confirmed.
12.4 Clearing Patient ID and patient prescription
The treatment parameters will revert to default/startup values if the Clear button is
pressed before the treatment starts. If the Clear button is pressed when the AK 98
dialysis machine is in treatment mode, only the patient's ID will be cleared. Parameter
settings will remain unchanged.