B:22 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
105 Blood pump door is not closed
105 Blood pump door is not closed
Close blood pump door.
Appears:
When the blood pump door is not properly closed during
treatment or priming.
Machine actions:
The blood pump stops.
Operator action:
1. Check that the blood pump segment is positioned
correctly.
2. Check that the blood pump rotor is positioned correctly.
3. Close the blood pump door.