AK 98 Dialysis Machine - Alarms B:43
1.3 Medium priority alarms
200 Arterial pressure limits too wide
200 Arterial pressure limits too wide
Centralize limits.
Appears:
When the arterial pressure alarm window is wider than the
preset value.
Machine actions:
ā— Alarm sound is delayed for 2 minutes.
ā— The arterial pressure control is activated.
Operator action:
1. Check that correct blood flow rate is set.
2. Press the arterial pressure control.
3. Press Center arterial.
4. Check that the arterial pressure limits are correct.