AK 98 Dialysis Machine - Alarms B:55
212 Low dialysis fluid temperature
212 Low dialysis fluid temperatureī¼
Appears:
When the dialysis fluid temperature has reached the set
temperature low alarm limit.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
1. Press the Fluid button.
2. Select the Temp tab to reach the temperature settings.
3. Check that the machine is connected to the power
socket.
4. Check the inlet water supply and the power supply.
Take appropriate actions when you find an issue.
5. Wait until the dialysis fluid temperature is correct. Then
the dialysis fluid automatically enters the dialyzer.
If necessary, call for an authorised technician or end the
treatment.