AK 98 Dialysis Machine - DIASCAN A:137
2) Select the Setup tab.
3) Press Interval.
4) Select 60 minutes , 30 minutes or
Single.
5) Press OK to start the measurement.
Results
The result of the measuring is displayed in the clearance measurement readout field.
The clearance measurement readout field shows:
ā— latest measured clearance value.
ā— time of latest measuring.
ā— current/estimated end of treatment Kt.
ā— the Diascan button (opens the Diascan menu).
9.4 Check Kt/V
The dialysis machine can calculate Kt/V as a single measurement or at intervals.
If Kt/V measurement is required, distribution volume has to be set before the
measurement check takes place. The distribution volume is patient-related (based on
the patient's dry weight) and has to be properly estimated and set by the operator in
order to obtain a correct Kt/V value.
Distribution volume is the urea distribution volume in litres (water in the body),
estimated for the individual patient, and based on the patient’s dry weight. The
estimation is based on certain formulae used for this purpose, chosen by the
prescribing physician responsible for the patient. The distribution volume has to be set
by the machine operator in order for a correct Kt/V value to be obtained.
NOTE!
To get a reliable Kt/V value, it is very important that the dialysis machine calculates
with an accurate distribution volume for the patient. Therefore fill in the patient
information with care in Distribution volume.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals