B:66 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
224 Second internal fluid leakage detected
224 Second internal fluid leakage detected
Check patient weight loss. Discontinue
treatment and call service technician.
Appears:
When the machine has detected a second fluid leakage
during treatment.
Machine actions:
ā— Close valves to stop leakage to the environment.
Operator action:
1. Check patient weight loss.
2. Discontinue treatment. Check patient and call service
technician
3. Call technical service.