B:98 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
523 Dialysis fluid flow is too low
523 Dialysis fluid flow is too low
Appears:
When the flow rate of the dialysis fluid is below the set low
alarm limit. This is because the flow rate is manually set
below the alarm limit or there is an obstruction in the
dialysis fluid flow path.
Machine actions:
None.
Operator action:
If possible, increase the dialysis fluid flow rate. If this is not
possible, contact an authorised technician.
If the attention is not solved, it will appear again in the next
treatment.
Mute time:
Permanent.