AK 98 Dialysis Machine - Attentions B:83
508 Blood pump door is not closed
508 Blood pump door is not closed
Close blood pump door and if not possible,
check the mounting of the pump rotor.
Appears:
When the blood pump door has not been properly closed.
Machine actions:
ā— The blood pump stops.
ā— After 2 minutes there is a buzzer sound.
Operator action:
Close the blood pump door.
Mute time:
2 minutes.