B:146 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
573 Low flow heat is stopped
573 Low flow heat is stoppedī¼
To continue to drain phase press Confirm.ī¼
Appears:
When the WRO 300 H cannot deliver hot water during
integrated heat disinfection program with WRO 300 H. The
low flow heat phase (where the machine receives hot water
from the WRO 300 H) will not start.
This attention can appear if you manually choose to start a
low flow heat phase but do not confirm it within 5 minutes.
Machine actions:
The low flow heat phase does not start.
Operator action:
1. Press Confirm to confirm the attention. The machine
now enters drain phase.
2. Contact an authorised technician.
Mute time:
Permanent.