HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Alarms B:19
102 Blood is detected during functional check
102 Blood is detected during functional checkī¼
Blood in priming detector. Functional check is
stopped.ī¼
Appears:
When the priming detector detects blood during functional
check.
Machine actions:
ā The blood pump stops.
ā The venous blood line clamp closes.
Operator action:
1. Make sure that the patient is not connected to the blood
lines.
2. Clean the priming detector lens.
If previous steps do not help, call an authorised technician.