A:150 AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
2) Select the Chemical tab.
3) Select disinfectant.
4) Connect the disinfectant. From the front of
the machine: connect the blue concentrate
connector to the yellow pick-up tube.
5) Place the pick-up tube in the disinfectant
solution.
6) Follow instructions on the screen.
WARNING!
After a chemical disinfection
program, a test for residues
shall be performed prior to
connecting a patient to the
dialysis machine to make
sure that no residues exist in
the dialysis machine.
10.4.3 Start a central chemical disinfection
In a central chemical disinfection the disinfectant comes from the central water
supply system.
This function has to be preset by an authorised service technician.
Procedure
1) Press the Disinfection button.
2) Select the Chemical tab.
3) Select the disinfectant that is set for central
administration.
4) When you get an attention that the dwell
time is over, and you have confirmed that
no disinfectant is circulated in the central
water system the machine starts a rinse
program.