A:130 AK 98 Dialysis Machine - Measuring blood pressure
HCEN12745 Revision 10.2016
Program version 2.xx
CAUTION!
If the patient has other monitoring devices on the same arm as the cuff, these
devices may not work as they should while the blood pressure is measured.
CAUTION!
Do not apply a cuff to a limb that has restricted blood flow as this may affect the
blood pressure measuring.
CAUTION!
Do not apply a cuff to a limb with an access, for example a fistula or a graft as this
may have a negative effect of the fistula or graft.
8.3 Direct blood pressure measuring
Procedure
1) Press the BPM readout field.
The measurement starts. The
heart in the read out field starts
flashing and BPM is flashing in the
status bar.
NOTE!
First inflation pressure is 180
mmHg.
Results
The measured value appears in the BPM read out field.
To stop measuring: Press the BPM read out field to stop an ongoing measurement.
8.4 Interval blood pressure measuring
Procedure
1) Press the BPM button.
2) Select the Setup tab.
3) Press Auto mode.
4) Press Interval time.
5) Set how often the blood pressure monitor
should measure blood pressure, a value
between 5 and 60 minutes.
6) Press OK.
7) Press Activate to activate interval blood
pressure measurement.