HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Disinfection and cleaning A:145
When disinfection is completed
ā— An operator message appears when the disinfection program is complete. To start
a new treatment: Press Confirm to reset previous values and start the functional
check.
NOTE!
You can stop a disinfection program at any time. The machine will interrupt the
procedure and finish off with a rinse and drain phase.
Stopping a disinfection program
ā— To stop a disinfection program: Press the Disinfection button.
ā— Select the tab with the actual disinfection program.
ā— Press Stop for the actual disinfection program.
Set the machine to switch off after completed program
Press the On/Off button for three seconds while a program is running to set the
dialysis machine to switch off after the program is completed.
ā— The machine can also be preset by an authorised service technician to
automatically switch off after a complete program.
10.3 Heat disinfection
10.3.1 Description of heat disinfection
During the heat disinfection program the inlet water is heated up and flushed through
the fluid monitor. The AK 98 dialysis machine fluid path is constructed in a way that
prevents possible contaminants in the post-dialyzer fluid path from reaching the
patient.
Heat disinfection reduces the number of viable microorganisms. The program can be
combined with cleaning and decalcification.
A heat disinfection begins with a rinse in order to rinse out possible residues from
concentrates used in the last treatment, followed by the circulation phase. The
circulation phase consists of a number of cycles , during each cycle heated water is
passed through all parts of the fluid path.
10.3.2 Cleaning and decalcification
Cleaning and decalcification are always done with a heat disinfection program.
Cleaning or decalcification solution is mixed with water, heated and flushed through
the fluid path during the heat disinfection program. A decalcification (CLEANCART C
cartridge or liquid citric acid) program cannot be replaced by a chemical disinfection
program.
The cleaning solution can be a CLEANCART A cartridge cleaning solution or sodium
hypochlorite. These alternatives remove fat, protein and organic material.
The decalcification solution can be a CLEANCART C cartridge solution or liquid citric
acid. These alternatives remove calcium-carbonate deposits.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals