B:82 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
507 Blood leak detector failure
507 Blood leak detector failureī¼
Clean detector and press Confirm.ī¼
Appears:
During functional check if the blood leak detector is dirty or
out of function.
Machine actions:
None.
Operator action:
1. Clean the blood leak detector, see Section 13.3 āClean
the blood leak detectorā on pageīA:173 in Operators
handbook.
2. Press Confirm to continue.
If the attention appears again, contact an authorised
technician.
Mute time:
Permanent.