B:100 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
525 Dialysis fluid not ready for treatment
525 Dialysis fluid not ready for treatment
Fluid is ready in {0} s.
Appears:
When ultrafilter flush is started.
Machine actions:
None.
Operator action:
1. Wait until the attention disappears and the bypass path
on the screen lights up green.
2. Disconnect the dialysis fluid tubes.
Mute time:
-