A:154 AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
5) Select Rinse.
6) Press OK.
7) Select day.
8) Set the time when the program should
finish.
9) Press OK.
10) To schedule more days in the week: repeat
steps 7-9.
10.5.3 To turn off rinse program for a certain day
Procedure
1) Press the Disinfection button.
2) Select the Tools tab.
3) Press Auto schedule.
4) Select day.
5) Press Off.
6) Press OK.
10.6 Machine storage with chemical disinfectant
10.6.1 Fill the dialysis machine with chemical disinfectant
If the dialysis machine is not to be used for more than 7 days it is recommended to fill
it with a chemical disinfectant before storage. It is recommended to use peracetic acid
as disinfectant during storage.
This function has to be preset by an authorised service technician.