AK 98 Dialysis Machine - Attentions B:149
576 Negative UF rate
576 Negative UF rateī¼
Check patient for weight gain. Check for
obstructions.ī¼
Appears:
When there is a backfiltration of 500 mL to the patient or a
backfiltration rate >50 mL/min for 5 minutes.
Machine actions:
None.
Operator action:
1. Check if there are obstructions in the outlet tube (the
tube where the used dialysis fluid flows out from the
dialysis machine).
2. Remove any obstructions and press Confirm.
3. Check that the UF rate is correct. If not, contact an
authorised technician.
Mute time:
2 minutes.
WARNING!
If no action is taken for this attention, the patient may get
incorrect weight loss during the treatment.