B:24 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
107 Blood pump is stopped for too long
107 Blood pump is stopped for too long
Appears:
When the blood pump has been stopped for a configurable
time.
Machine actions:
None.
Operator action:
Press the Blood pump button to start the blood pump
when the patient is connected and the dialysis machine
detects blood. Make sure that the cause of the alarm has
been corrected.