B:120 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
546 Functional check is on hold
546 Functional check is on holdī¼
Water temperature is too high.ī¼
Appears:
When the inlet water temperature is too high during
functional check.
Machine actions:
The functional check stops. The functional check will
automatically continue when the inlet water temperature
has decreased.
Operator action:
Check possible cause:
ā The inlet water temperature has not decreased below
45 °C within 5 minutes after the high temperature test is
started.
ā The inlet water temperature has not decreased below
39 °C within 5 minutes after the high temperature test is
finished.
If the attention does not disappear, contact an authorised
technician.
Mute time:
Permanent.