A:132 AK 98 Dialysis Machine - Measuring blood pressure
HCEN12745 Revision 10.2016
Program version 2.xx
An alarm will occur if a value is outside the set alarm limits.
Procedure
1) Press the BPM button.
2) Select the Alarm tab.
3) Press Set limits.
4) Select the alarm limits you want to set.
Enter the alarm limit value.
5) Press OK.
6) Repeat step 4 and 5 for all of the alarm
limits you want to set. Close the Set
Limits menu.
7) Press the activate button for alarm limits to
activate the new settings.
8.7 Patient care during blood pressure measuring
8.7.1 All patients
CAUTION!
Check the material used in cuff before use to avoid allergic symptoms. Allergic
symptoms caused by the fabric material of the cuff could lead to exanthema and
urticaria with swollen skin and intense itching in the cuff area.
CAUTION!
Check the material used in cuff before use to avoid allergic symptoms. Multiple or
few petechia (small purple spots containing blood) on the forearm following the
application of the cuff which may lead to decrease of thrombocytes in the blood.
Inflammation of a vein (phlebitis) may be observed.
CAUTION!
Continuous cuff pressure due to kinks in the connection tubing may result in blood
flow interference and cause harmful injury to the patient.
CAUTION!
Too frequent blood pressure measurements may cause injury to the patient due to
blood flow interference.