B:50 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
208 Incorrect dialysis fluid composition
208 Incorrect dialysis fluid compositionī¼
Check set values and connected concentrates.ī¼
Appears:
When the dialysis fluid composition deviation (the relation
between the acidic and the bicarbonate concentrate) is too
high.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
1. Check that the concentrate container is filled.
2. Check what concentrate is set in the dialysis machine.
a) Press the Fluid button.
b) Select the Conc tab.
3. Check that the concentrate that is set is the same as
the concentrate connected.
4. Check that the proper concentrate has been connected
correctly.
5. Check that there is no air in the concentrate tube.