B:20 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
103 Blood is detected in assisted priming
103 Blood is detected in assisted priming
Blood in priming detector. Blood pump is
stopped.
Appears:
When the priming detector detects blood during assisted
priming procedure.
Machine actions:
The blood pump stops.
Operator action:
1. Make sure that the patient is not connected to the blood
lines.
2. Clean the priming detector lens.
If previous steps do not help, call an authorised technician.