AK 98 Dialysis Machine - Alarms B:27
110 High diastolic blood pressure
110 High diastolic blood pressure
To clear the alarm press Confirm.
Appears:
When the patient's diastolic blood pressure is outside the
set high alarm limit.
Machine actions:
None.
Operator action:
1. Check the blood pressure history:
- Press the BPM button.
- Select the History tab.
- Press History.
- Observe the measurement results from the treatment.
2. Do standard patient checks and take measures for
patients with circulatory disturbances.
3. If desired, measure the blood pressure manually (see
Section 8.3 ā€œDirect blood pressure measuringā€ on
pageī™›A:130 in Operators handbook).
4. If possible, adjust the alarm limits to the patient’s status:
- Press the BPM button.
- Select the Alarm tab.
- Press Set limits.
- Press Diastolic High and adjust the alarm limit.
5. Press OK.