B:78 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
503 Battery failure
503 Battery failureī¼
Blood pump will not be able to run in case of
power failure. To continue, change battery or
press Confirm.ī¼
Appears:
ā When the backup battery does not pass functional
check.
ā After treatment if the backup battery test fails.
If the functional check fails, the alarm sound will be delayed
for 1 minute.
Machine actions:
None.
Operator action:
Press Confirm.
Mute time:
Permanent.