B:168 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
599 UF rate lower than minimum set
599 UF rate lower than minimum set
Calculated UF rate lower than set min UF rate.
Adjust time and/or UF volume or min UF rate.
Appears:
When the machine needs to decrease the UF rate below
the set minimum UF rate in order to obtain the set UF
volume at the end of treatment.
Machine actions:
Alarm sound is delayed for 2 minutes.
Operator action:
Change one or several of these settings:
ā— Decrease the set minimum UF rate
ā— Decrease treatment time to increase the UF rate
ā— Increase UF volume
Mute time:
2 minutes.