AK 98 Dialysis Machine - Attentions B:153
580 No backfiltration warning
580 No backfiltration warningī¼
TMP alarm limit/s below 0 mmHg. To accept and
continue treatment press Confirm.ī¼
Appears:
When the low TMP alarm limit is set below zero.
The appearance of the attention can be preset.
Machine actions:
None.
Operator action:
Since the TMP alarm limits have been set below zero there
may be backfiltration to the patient without TMP alarm.
Mute time:
-