B:48 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
206 Incorrect acidic concentrate
206 Incorrect acidic concentrateī¼
Check set value and connected concentrate.ī¼
Appears:
When the selected concentrate does not match the
concentrate connected to the machine.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
1. Check that the concentrate container is filled.
2. Check that there is no air in the concentrate tube.
3. Check what concentrate is set in the dialysis machine.
a) Press the Fluid button.
b) Select the Conc tab.
4. Check that the concentrate that is set is the same as
the concentrate connected.
5. Check that the proper concentrate has been connected
correctly.