B:112 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
538 Fluid tube sensor test
538 Fluid tube sensor testī¼
Attach fluid tubes to safety couplings. To
interrupt test press Confirm.ī¼
Appears:
When the dialysis machine cannot detect that the dialysis
fluid tubes are attached to the safety couplings. The
corresponding machine test cannot be made.
Machine actions:
The ongoing disinfection or rinse program stops.
Operator action:
To continue the disinfection or rinse program:
1. Check that the dialysis fluid tubes are properly attached
to the safety couplings.
2. Press Confirm. The disinfection or rinse program
continues.
As an alternative, consult an authorised technician.
Mute time:
2 minutes.