EasyManuals Logo

Gambro AK 98 Operator's Manual

Gambro AK 98
386 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Disinfection and cleaning A:149
3) Press Auto schedule.
4) Select day.
5) Press Off.
6) Press OK.
10.3.8 Integrated heat disinfection with a WRO 300 H unit
10.3.8.1 Integrated heat disinfection with a WRO 300 H unit
An integrated heat disinfection program has an extra phase compared to normal heat
disinfection. In this phase hot water from the water purification unit WRO 300 H slowly
flows through the dialysis machine. This means that also the inlet water tube is
disinfected. Therefore, choose integrated heat disinfection if possible. For more
information, see Section 11 ā€œDisinfection with the AK 98 dialysis machine and WRO
systemā€ on pageī™›A:159
10.4 Chemical disinfection
10.4.1 About chemical disinfection
During the chemical disinfection program the machine is filled with a concentrated
disinfectant . Concentrated disinfectant is mixed with water to the correct
concentration in the machine. The diluted solution fills all parts of the fluid path. After a
certain time (dwell time) the fluid path is rinsed and drained.
Heat disinfection can be replaced by chemical disinfection. But note that decalcification
(CLEANCART C cartridge or liquid citric acid) cannot be replaced by chemical
disinfection.
10.4.2 Start a chemical disinfection
Procedure
1) Press the Disinfection button.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals