AK 98 Dialysis Machine - Alarms B:65
222 Venous pressure limits too wide
222 Venous pressure limits too wideī¼
Centralize limits.ī¼
Appears:
When the venous pressure alarm window is wider than the
preset value.
Machine actions:
ā Alarm sound is delayed for 2 minutes.
ā The venous pressure control is activated.
Operator action:
1. Check that correct blood flow rate is set.
2. Press the venous pressure control.
3. Press Center venous.
4. Check that the venous pressure limits are correct.