B:144 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
570 Kt/V target will not be reached
570 Kt/V target will not be reached
Check treatment settings and check/adjust Kt/V
alarm limit. To continue press Confirm.
Appears:
Kt/V forecast value is below the set alarm limit for Kt/V.
Machine actions:
None.
Operator action:
Check for a cause, adjust treatment parameters or adjust
the alarm limit for Kt/V.
Mute time:
Permanent.