AK 98 Dialysis Machine - Isolated ultrafiltration A:115
6.2.2 How to add a second and subsequent phase of isolated UF
Second and subsequent phases of isolated UF can be activated at any point during
the treatment.
The time and UF volume for the new phase must be added to the previous isolated UF
time and UF volume to create a cumulative amount.
Example: If the 1st phase was set to 30 minutes and 0.5L and a second phase is to be
set with the same values, the settings need to be 60 minutes and 1.0L which is the
total of the two phases together.
The values set for the second and subsequent phases must always be larger than the
accumulated isolated time and UF volume.
6.2.3 Deactivate isolated ultrafiltration
If necessary, isolated UF can be deactivated at any time during the phase.
Procedure
1) Press the Fluid button.
2) Press Isolated UF.
3) Deactivate Isolated UF.
6.3 Additional information
6.3.1 Heparin
If the heparin pump is used, it will run during the complete isolated UF, unless isolated
UF is performed after the diffusion phase. Settings for Heparin bolus volume,
Heparin flow rate and Stop time are kept.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals