AK 98 Dialysis Machine - Attentions B:173
604 Venous clamp test failed
604 Venous clamp test failedī¼
Check venous line placement in venous clamp.
To continue press Confirm.ī¼
Appears:
When the venous clamp test fails.
Machine actions:
The machine stops the venous clamp test until the attention
is confirmed.
Operator action:
Check that the blood lines are correctly inserted into the
venous clamp and confirm the attention.
Mute time:
2 minutes.