AK 98 Dialysis Machine - Alarms B:67
225 Internal fluid leakage detected
225 Internal fluid leakage detected
Check patient weight loss. To continue, wipe
fluid detector container dry.
Appears:
When the machine has detected a fluid leakage during
treatment.
Machine actions:
ā— Close valves to stop leakage to the environment.
Operator action:
1. Check patient weight loss.
2. Wipe the fluid detector tray dry.
3. Continue treatment.