AK 98 Dialysis Machine - Attentions B:145
571 Leakage test failed
571 Leakage test failed
Check fluid tubes. To continue press Confirm.
Appears:
When the fluid tubes are not properly attached to the safety
couplings.
Machine actions:
The leakage test is repeated until it is approved, or until the
function check is stopped.
Operator action:
Check the fluid tubes and confirm the attention.
Mute time:
2 minutes.