A:186 AK 98 Dialysis Machine - Technical data and specifications
HCEN12745 Revision 10.2016
Program version 2.xx
14.2.3 Extracorporeal blood loss due to coagulation
Detection method Supervision of the stop time of the blood pump. The default alarm limit
is 60 seconds.
Alarm The alarm
107 Blood pump is stopped for too longī¼
14.2.4 Dialysis fluid preparation
High temperature
alarm (fixed)
40 °C (±0.5 °C)
Low temperature
alarm (fixed)
32.5 °C (±0.5 °C)
Table 14-20. Temperature
Alarm limits Error in the preparation of the dialysis fluid may lead to conductivity
alarms. The maximum possible deviation of each ion before a
conductivity alarm is issued, is ±5 % for sodium ion and ±25 % for
bicarbonate ion.
Accuracy 0.2 mS/cm
Table 14-21. Conductivity
14.2.5 TMP
TMP TMP is defined as the difference, P
b out
- P
d out
, where P
b out
is the
venous pressure and P
d out
is the pressure measured in the dialysis
fluid, where it enters the machine after the dialyzer. The displayed TMP
value is compensated for vertical difference between the measure
points.
Alarm limits User settable between -200 to 500 mmHg
Accuracy ±10 mmHg or ±6 %, whichever is largest (within range ±500 mmHg)
14.2.6 Blood leakage detection
Detection method Infrared light detector.
Sensitivity Alarm will be given if ā„0.35 mL/min blood with hematocrit at 32 % is
entered into the dialysis fluid flow.
14.3 Alarm sound pressure
14.3.1 Alarm sound pressure
The alarm sound pressure is presetable to minimum 55 dB (A) and maximum 85 dB
(A). The default preset is 65 dB (A).
14.4 Physical data
14.4.1 Dimensions and weight
Width; machine Approx. 345 mm
Width; stand Approx. 585 mm