HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions B:143
569 Kt/V target reached
569 Kt/V target reachedī¼
To continue press Confirm.ī¼
Appears:
When the set target value for Kt/V is reached.
Machine actions:
None.
Operator action:
Press Confirm to confirm the attention.
It is possible to increase the Kt/V target value. The machine
will recalculate time to target.
Mute time:
-