A:96 AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
CAUTION!
It is important to set a correct ultrafiltration rate for the patient, to avoid incorrect fluid
removal. Set a minimum ultrafiltration rate according to the instructions for the
specific dialyzer.
CAUTION!
Once treatment parameters are set, you must verify that all calculated treatment
values are as intended.
CAUTION!
Additional measures to supervise the patient weight loss is recommended when
treating low weight patients or when performing long treatments. For Ultrafiltration
details, refer to Technical data and specifications.
Procedure
1) Press the Time indicator.
2) Press UF volume.
3) Calculate the patientās ultrafiltration
volume, see Section 4.2.10 āSet
ultrafiltration volumeā on pageīA:95.
4) Enter the patientās ultrafiltration volume.
5) Press OK.
4.2.11 Set heparin values
Before you begin
WARNING!
Make sure to set the prescribed heparin values. Incorrect heparin settings may
cause blood coagulation or internal bleeding.
Procedure
1) Press the Blood button.
2) Select the Heparin tab.