0: Common commands Model DMM7510 7½ Digit Graphical Sampling Multimeter
8 DMM7510-901-01 Rev. B / May 2015
*SRE
This command sets or clears the bits of the Service Request Enable Register.
Type Affected by Where saved Default value
Command and query
:STATus:PRESet
Not applicable 0
Usage
*SRE <n>
*SRE?
Clear the Status Request Enable Register: 0
Set the instrument for an SRQ interrupt: 32
Details
This command sets or clears the individual bits of the Status Request Enable Register.
The Status Request Enable Register is cleared when power is cycled or when a parameter value of 0
is sent with this command.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register.
Bit
Decimal
value
Constants When set, indicates the following has occurred:
An enabled event in the Measurement Event Register has
occurred.
An error or status message is present in the Error Queue.
3 8
An enabled event in the Questionable Status Register has
A response message is present in the Output Queue.
An enabled event in the Standard Event Status Register
has occurred.
6 64 Not used Not used.
7 128
An enabled event in the Operation Status Register has
Constants are only available if you are using the TSP command set. If you are using the SCPI
command set, you must use the decimal values.
Example
Clear the bits of the Status Request Enable Register.
Also see
Understanding bit settings (on page 14)