6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-222 DMM7510-901-01 Rev. B / May 2015
:TRIGger:LOAD "Empty"
This command resets the trigger model.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:LOAD "Empty"
Details
When you load this predefined trigger model, any blocks that have been defined in the trigger model
are cleared so the trigger model has no blocks defined.
Example
TRIG:BLOC:BUFF:CLEAR 1
TRIG:BLOC:MEAS 2
TRIG:BLOC:BRAN:COUN 3, 5, 2
TRIG:BLOC:DEL:CONS 4, 1
TRIG:BLOC:BRAN:COUN 5, 3, 2
Reset trigger model settings.
Clear defbuffer1
at the beginning of execution
of the trigger model.
Loop and take 5 readings.
Delay 1 s.
Loop three more times back to block 2.
At the end of execution, 15 readings are stored
in
.
Also see
None