8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-138 DMM7510-901-01 Rev. B / May 2015
dmm.measure.bias.level
This attribute selects the amount of current the instrument sources when it makes measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
1 mA
Usage
value = dmm.measure.bias.level
dmm.measure.bias.level = value
10 µA, 100 µA, 1 mA
Functions
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Details
Selects the amount of current that is sourced by the instrument to make measurements.
Example
dmm.measure.bias.level = 0.0001
µ
Also see
dmm.measure.bias.actual (on page 8-137)
dmm.measure.configlist.catalog()
This function returns the name of one measure configuration list that is stored on the instrument.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
dmm.measure.configlist.catalog()
Details
You can use this command to retrieve the names of measure configuration lists that are stored in the
instrument.
This command returns one name each time you send it. This command returns nil to indicate that
there are no more names to return. If the command returns nil the first time you send it, no measure
configuration lists have been created for the instrument.