Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6: SCPI command refere
nce
DMM7510-901-01 Rev. B / May 2015 6-109
[:SENSe[1]]:<function>:THReshold:LEVel
This command determines the signal level where the instrument makes frequency or period measurements.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:THReshold:LEVel <n>
[:SENSe[1]]:<function>:THReshold:LEVel DEFault
[:SENSe[1]]:<function>:THReshold:LEVel MINimum
[:SENSe[1]]:<function>:THReshold:LEVel MAXimum
[:SENSe[1]]:<function>:THReshold:LEVel?
[:SENSe[1]]:<function>:THReshold:LEVel? DEFault
[:SENSe[1]]:<function>:THReshold:LEVel? MINimum
[:SENSe[1]]:<function>:THReshold:LEVel? MAXimum
<function>
The function to which the setting applies; see Functions
<n>
The level:
−
700 V to 700 V; dependent on range
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
You need to set an appropriate voltage trigger level in order for the frequency counter to operate
properly. The frequency counter only counts cycles when the signal amplitude reaches the trigger
level. For example, if you set the trigger level for 10 V, any cycles with peak amplitude less than 10 V
are not counted.
You must select a specific threshold range (autorange must be set to off) before setting a level that is
not zero.
Example
:FUNC "FREQ"
:FREQ:THR:RANG 10
:FREQ:THR:LEV 5
Set the measure function to frequency.
Set the threshold range to 10 V.
Set the threshold level to 5 V.
Also see
[:SENSe[1]]:<function>:THReshold:RANGe (on page 6-110)
[:SENSe[1]]:<function>:THReshold:RANGe:AUTO (on page 6-111)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals