8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-254 DMM7510-901-01 Rev. B / May 2015
trigger.blender[N].clear()
This function clears the blender event detector and resets the overrun indicator of blender N.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.blender[N].clear()
The blender number (1 or 2)
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
trigger.blender[2].clear()
Clears the event detector for blender 2.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Configuration script false (AND)
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
• true: OR operation
•
: AND operation
The blender number (1 or 2)
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected
events (AND) before signaling an output event.