[:SENSe[1]]:<function>:RELative:STATe
This command enables or disables the application of a relative offset value to the measurement.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:RELative:STATe <state>
[:SENSe[1]]:<function>:RELative:STATe?
The function to which the setting applies; see Functions
Disable the relative offset: OFF or 0
Enable the relative offset: ON or 1
Functions
Details
When relative measurements are enabled, all subsequent measured readings are offset by the
relative offset value. You can enter a relative offset value or have the instrument acquire a relative
offset value.
Each returned measured relative reading is the result of the following calculation:
Displayed reading = Actual measured reading - Relative offset value
Example
:SENS:VOLT:REL 5
Set the measurement function to volts with a
relative offset of 5 V and enable the relative
offset function.
Also see
Relative offset (on page 3-4)
[:SENSe[1]]:<function>:RELative (on page 6-92)
[:SENSe[1]]:<function>:RELative:ACQuire (on page 6-94)