7½ Digit Graphical Sampling Multimeter Reference Manual Section 6: S
DMM7510-901-01 Rev. B / May 2015 6-67
[:SENSe[1]]:<function>:AVERage:COUNt
This command sets the number of measurements that are averaged when filtering is enabled.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
10
Usage
[:SENSe[1]]:<function>:AVERage:COUNt <n>
[:SENSe[1]]:<function>:AVERage:COUNt DEFault
[:SENSe[1]]:<function>:AVERage:COUNt MINimum
[:SENSe[1]]:<function>:AVERage:COUNt MAXimum
[:SENSe[1]]:<function>:AVERage:COUNt?
[:SENSe[1]]:<function>:AVERage:COUNt? DEFault
[:SENSe[1]]:<function>:AVERage:COUNt? MINimum
[:SENSe[1]]:<function>:AVERage:COUNt? MAXimum
The function to which the setting applies; see Functions
The number of readings required for each filtered measurement (1 to 100)
Functions
Details
The filter count is the number of readings that are acquired and stored in the filter stack for the
averaging calculation. When the filter count is larger, more filtering is done and the data is less noisy.
Example 1
CURR:AVER:TCON MOV
CURR:AVER ON
For current measurements, set the averaging
filter type to moving average, with a filter count of
10.
Enable the averaging filter.
Example 2
RES:AVER:TCON MOV
RES:AVER ON
For resistance measurements, set the averaging
filter type to movin
g average, with a filter count of
10.
Enable the averaging filter.